Product Description
The Tektronix TDS5104B Digital Oscilloscope offers the industries first easily customizable oscilloscope user interface. MyScope control windows is a revolutionary new feature that allows you to build your own control windows with only the controls, features, and capabilities that you care about and are important in your job. For the first time, you can pull all the functionality you need from all the various parts of the oscilloscope into one control window, effectively creating your own personalized toolbox of oscilloscope features.
No longer do you need to search through menus for features or re-learn how to drive the oscilloscope after a break from the lab. MyScope control windows enable you to spend your valuable time focused on the task at hand rather than navigating menus on your oscilloscope. And creating these custom control windows isnt a long drawn-out or complex process. They are easily created in a matter of minutes using a simple, visual, drag and drop process. Once created, these custom control windows are easily accessed through a dedicated MyScope button and menu selection on the oscilloscope button/menu bar, just like any other control window.
You can make an unlimited number of custom control windows, enabling each person who uses the oscilloscope, in a shared environment, to have their own unique control window. Since the control windows are stored as files on the hard drive, they can easily be transferred to other TDS5104B Series oscilloscopes, or they can even be e-mailed to a co-worker around the world when the need arises. MyScope control windows will benefit all oscilloscope users, from eliminating the ramp-up time that many face when returning to the lab after not using an oscilloscope for a while, to the power user who can now operate far more efficiently. Everything you need is found in one control window rather than having to constantly navigate through menu after menu to repeat similar tasks.
1 GHz Bandwidth Model
4 Channel Models
5 GS/s Sample Rate
Up to 16 M Record Length
100,000 wfms/s Maximum Waveform Capture Rate
MyScope Custom Control
Windows Enhance Productivity
Right Mouse Click Menus for Exceptional Efficiency
OpenChoice® Platform with Windows 2000 Delivers Built-in Networking and Analysis
Small Footprint/Light Weight
10.4 in. Bright Display
Suite of Advanced Triggers
Communication Mask Testing
Pass/Fail Limit Testing
Remote Viewing and Control
E-mail on Event
CD-RW Drive
Interoperability with Tektronix Logic Analyzers
GPIB Controller
Applications:
Digital Design and Debug
Mask Testing for Telecomm/Datacomm/Video Standards
Investigation of Transient Phenomena
Power Measurements
Video Design and Debug
DVD, Jitter and Timing, Spectral Analysis
Automotive Electronics
Manufacturing Test
Electro-mechanical
Bio-medical
Industrial Control
Specifications:
Vertical System | |
---|---|
Input Channels | 4 |
Analog Bandwidth (-3 dB) 5 mV/div - 1 V/div | 1 GHz |
Calculated Rise Time 5 mV/div (typical) | 300 ps |
Hardware Bandwidth Limits | 150 MHz or 20 MHz |
Input Coupling | AC, DC, GND |
Input Impedance, 1 MΩ | ±1% |
Input Impedance, 50 Ω | ±2.5% |
Input Sensitivity, 1 MΩ | 1 mV/div to 10 V/div |
Input Sensitivity, 50 Ω | 1 mV/div to 1 V/div |
Vertical Resolution | 8-bits (>11-bits w/ averaging) |
Max Input Voltage, 1 MΩ | 150 V CAT I, ≤ 400 V peak. Derate at 20 dB/decade to 9 VRMS above 200 kHz |
Max Input Voltage, 50 Ω | <100 mV/div <1 VRMS |
DC Gain Accuracy | 1.5% with offset set to 0 V |
Offset Range, 1 MΩ | 1 mV/div - 99.5 mV/div ±1 V |
Offset Range, 50 Ω | 1 mV/div - 50 mV/div ±0.5 V |
Channel-to-channel Isolation for Any Two Channels at Equal Vertical Scale | ≥100:1 at ≤100 MHz and ≥30:1 at >100 MHz up to the rated bandwidth |
Timebase System | |
---|---|
Timebase Range | 200 ps/div to 1000 s/div |
Timebase Delay Time Range | (s/div " 10) to 1000 s |
Channel-to-channel Deskew Range | ±75 ns |
Timebase Accuracy | 15 ppm |
Delta Time Measurement Accuracy | (0.06/sample rate + 15 ppm " |Reading|)RMS |
Trigger Jitter (RMS) | 8 psRMS (typical) |
Long Term Sample Rate and Delay Time Accuracy | ±15 ppm over any ≥1 ms interval |