Microplate Reader Thermo Multiskan Ascent 96 & 384 by AnytimeLabTrader, LLC
- Item Condition:
- Brand:
- Model#:
- Location:
- Price:
Product Description
Microplate Reader Thermo Multiskan Ascent
Current VWR List Price: 8,001.00
This is a Nice PreOwned Microplate Reader (Reads Both 96 & 384 Plates) by Thermo Labsystems Model #354 and Model Name Multiskan Ascent. Unit is in Excellent Working and Physical Condition. Unit Was Tested Against a Calibration Plate and all Values Were Within 2% of Target, Results Provided Below. Comes Complete with Dell Computer (Running Ascent Software), LCD, Keyboard, Mouse, User's Manual, All Communication and Power Cables. Unit is Ready to Use Within Minutes of Receiving. Reader has 405, 450, 492, 570, 595, and 650nm Filters Installed. Unti will be Guaranteed for 30 Days Allowing Enough Time for Your Lab To Fully Test and Inspect.
Features:
- Wavelength range :340-850 nm (With Appriopriate Filters)
- Plate Formats: 96- and 384-well plates
- Measurement Types: Single, Dual Wavelength, Kinetic, or Dual Kinetic
- Measurement Modes: Continous or Stepping
- Linearity (96-well plate) :0-4 Abs, +/-2% 0-3 Abs, +/-2%, 340 nm and 405 nm
- Linearity (384-well plate) :0-3 Abs, +/-2% 0-2 Abs, +/-2% with continuous mode
- Read-out range :0-6 Abs
- Accuracy :+/-1% or 0.003 Abs, whichever is greater (0-2 Abs) +/-2% (2-3 Abs)
- Precision :CV<0.2% (0.3-3 Abs), CV<1.0% (3-4 Abs) stepping mode
- Measurement time (96-well plate) :9 s, continuous 14 s, stepping
- Measurement time (384-well plate) :14 s, continuous 42 s, stepping
- Shaking :Linear shaking, 3 speeds
- Filters: Eight-position filter wheel with 405, 450, 490 and 650nm Filters!
Calibration Test Plate Results:
1) 405nm Filter: * Expected .590, 1.250, 1.852 * Actual : .595, 1.254, 1.855
2) 450nm Filter: * Expected .548, 1.157, 1.714 * Actual: .554, 1.157, 1.710
3) 490nm Filter: * Expected .545, 1.150, 1.705 * Actual: .552, 1.158, 1.712
4) 650nm Filter: * Expected .537, 1.132, 1.675 * Actual: .539, 1.137, 1.679
If This Unit Does Not Meet Your Needs, Please See: Other Microplate Reader Available