  • Product Overview

The PS 501 is a contant-voltage, current-limiting, 0 to 20 Vdc power supply designed to operate in any compartment of a TM 500 Series power module. The PS 501 is a floating power supply in which either terminal may be grounded or floated to 350 Vdc + peak ac. Front panel controls provide continuously variable voltage selection as well as current limiting from 0 to at least 400 mA.

Several PS 501s or other suitable supplies can be combined for applications involving complementary circuitry, linear integrated circuits, or any circuitry requiring cascaded DC levels.

A ground-referenced +5 V auxiliary supply is included.

  • 20 V floating supply:
    • Voltage range: 0 V to at least 20 V
    • Rated output current: 400 mA
    • Current limit range: ≤10 mA to at least 400 mA
    • Maximum floating voltage: 350 Vdc + peak ac
    • Load effect (regulation): ±1 mV for a 400 mA load
    • Source effect (line regulation): ±5 mV for a ±10% line voltage change
    • Stability: Typically less than 0.1% +5 mV for 8 hours 
    • Ripple and noise: ≤0.5 mVp-p with 400 mA load
    • Transient recovery time: 20 µs or less to recover within 20 mV of final output voltage after a 400 mA change in output current
  • 5 V floating supply
    • Output voltage: +4.75 V to +5.25 V ground referenced
    • Rated output current: 1 A
    • Current limit: Non-adjustable and typically 1.5 A to 3 A
    • Load effect (regulation): ±100 mV for a 1 A load change
    • Source effect (line regulation): ±50 mV for a ±10% line voltage change
    • Stability: Typically less than 30 mV (0.5%) for 8 hours
    • Ripple and noise: ≤5 mVp-p with 1 A load