  • Product Overview

The MDO4000 delivers the functionality of an oscilloscope and a spectrum analyzer in a single instrument. The new MDO4000 Mixed Domain Oscilloscope Series gives engineers the unique ability to capture time-correlated analog, digital, and RF signals for a complete system view, helping them to rapidly solve complicated design issues.

More than 60 percent of oscilloscope users also use a spectrum analyzer to troubleshoot embedded system designs with integrated wireless functionality, requiring them to work in both the time and frequency domain. Traditionally, an engineer was either a mixed signal/digital engineer or an RF engineer. But with wireless becoming more commonplace, design engineers must often work in both domains. The MDO4000 Mixed Domain Oscilloscope Series is the first oscilloscope ever to integrate the functionality of a spectrum analyzer to provide a unique toolset which will save days or even weeks of debug time.

Based on the industry-standard MSO4000B Oscilloscope Series, you can now use your tool of choice, the oscilloscope, to look at the frequency domain rather than having to find and re-learn a spectrum analyzer. However, the power of the MDO goes well beyond simply observing the frequency domain as you would on a spectrum analyzer. The real power is in its ability to correlate events in the frequency domain with the time domain phenomena that caused them.

When both the RF channel and any analog or digital channels are on, the oscilloscope display is split into two views. The upper half of the display is a traditional oscilloscope view of the Time Domain. The lower half of the display is a Frequency Domain view of the RF input. Note that the Frequency Domain view is not simply an FFT of the analog or digital channels in the instrument, but is the spectrum acquired from the RF input.

The spectrum shown in the Frequency Domain view is taken from the period of time indicated by a short orange bar in the time domain view – known as the Spectrum Time. With the MDO4000 Series, Spectrum Time can be moved through the acquisition to investigate how the RF spectrum changes over time. And this can be done while the oscilloscope is live and running or on a stopped acquisition.
The upper half of the MDO4000 Series display shows the Time Domain view of the analog and digital channels, while the lower half shows the Frequency Domain view of the RF channel. The orange bar – Spectrum Time – shows the period of time used to calculate the RF spectrum.

  • 4 Analog Channels: 500 MHz or 1 GHz Bandwidth Models
  • 16 Digital Channels: MagniVu™ High-speed Acquisition provides 60.6 ps Fine Timing Resolution
  • 1 RF Channel
    • 50 kHz - 3 GHz or 50 kHz - 6 GHz Frequency Range Models
    • Ultra-wide Capture Bandwidth =1 GHz
  • Standard Passive Voltage Probes 3.9 pF Capacitive Loading and 500 MHz or 1 GHz Analog Bandwidth
  • Mixed Domain Analysis
    • Time-correlated Analog, Digital, and RF Signal Acquisitions in a Single Instrument
    • Wave Inspector® Controls provide Easy Navigation of Time-correlated Data from both the Time and Frequency Domains
    • Amplitude, Frequency, and Phase versus Time Waveforms derived from RF Input
    • Selectable Spectrum Time to Discover and Analyze how RF Spectrum Changes over Time – Even on a Stopped Acquisition
  • Spectral Analysis
    • Dedicated Front-panel Controls for Commonly Performed Tasks
    • Automated Peak Markers Identify Frequency and Amplitude of Spectrum Peaks
    • Manual Markers enable Non-peak Measurements
    • Trace Types Include: Normal, Average, Max Hold, and Min Hold
    • Detection Types Include: +Peak, –Peak, Average, and Sample
    • Spectrogram Display enables Easy Observation and Insight into Slowly Changing RF Phenomena
    • Automated Measurements Include: Channel Power, Adjacent Channel Power Ratio (ACPR), and Occupied Bandwidth (OBW)
    • Trigger on RF Power Level
    • Triggered or Free Run Spectral Analysis
  • Connectivity
    • Two USB 2.0 Host Ports on the Front Panel and Two on the Rear Panel for Quick and Easy Data Storage, Printing, and Connecting a USB Keyboard
    • USB 2.0 Device Port on Rear Panel for Easy Connection to a PC or Direct Printing to a PictBridge®-compatible Printer
    • Integrated 10/100/1000BASE-T Ethernet Port for Network Connection and Video Out Port to Export the Oscilloscope Display to a Monitor or Projector
  • Optional Serial Triggering and Analysis: Serial Protocol Trigger, Decode, and Search for I2C, SPI, USB, Ethernet, CAN, LIN, FlexRay, RS-232 / 422 / 485 / UART, MIL-STD-1553, and I2S / LJ / RJ / TDM
  • Optional Application Support
    • Advanced RF Triggering
    • Power Analysis
    • Limit and Mask Testing
    • HDTV and Custom Video Analysis
  • 10.4 inches (264 mm) Bright XGA Color Display
  • Small Footprint and Lightweight – Only 5.8 inches (147 mm) Deep and 11 lb (5 kg)

The MDO4104 is available in two models:

  • MDO4104-3: Mixed Domain Oscilloscope with four 1 GHz analog channels, sixteen digital channels, and one 3 GHz RF input
  • MDO4104-6: Mixed Domain Oscilloscope with four 1 GHz analog channels, sixteen digital channels, and one 6 GHz RF input

MDO4104 specifications:

  • Analog input channels: 4
    • Bandwidth: 1 GHz 
    • Calculated rise time, 5 mV/div (typical): 350 ps
    • Input impedance: 1 M? ±1%, 50 ? ±1%
    • Vertical resolution: 8 bits (11 bits with Hi Res)
    • Max input voltage, 1 M?: 300 VRMS CAT II with peaks =±425 V
    • Max input voltage, 50 ?: 5 VRMS with peaks <±20 V
    • Sample rate: 5 GS/s on 1-2 ch, 2.5 GS/s on 3-4 ch 
    • Maximum record length (all channels): 20 Mpoints
    • Time base range: 400 ps to 1000 s
    • Time base delay time range: -10 divisions to 5000 s
    • Channel-to-channel deskew range: ±125 ns
  • Digital input channels: 16
    • Threshold selections: TTL, CMOS, ECL, PECL, User Defined
    • User-defined threshold range: ±40 V
    • Maximum input voltage: ±42 Vpeak
    • Input dynamic range: 30 Vp-p =200 MHz, 10 Vp-p >200 MHz
    • Input impedance: 100 k?
    • Vertical resolution: 1 bit
    • Maximum sample rate (Main): 500 MS/s (2 ns resolution)
    • Maximum record length (Main): 20 Mpoints
    • Minimum detectable pulse width (typical): 1 ns
    • Channel-to-channel skew (typical): 200 ps
  • RF channels: 1
    • Frequency range: 50 kHz - 3 GHz for the MDO4104-3, 50 kHz - 6 GHz for the MDO4104-6
    • Real-time capture bandwidth: =1 GHz
    • Resolution bandwidth: 20 Hz - 10 MHz
    • Vertical units: dBm, dBmV, dBµV, dBµW, dBmA, dBµA
    • DANL: 50 kHz - 5 MHz: <–130 dBm/Hz (<–134 dBm/Hz typical); 5 MHz - 3 GHz: <–148 dBm/Hz (<–152 dBm/Hz typical); 3 GHz - 6 GHz: <–140 dBm/Hz (<–143 dBm/Hz typical)
    • 2nd and 3rd Harmonic Distortion (>30 MHz): <–55 dBc (<–60 dBc typical)
    • Average Continuous Power: +30 dBm (1 W)
    • Power level trigger amplitude range: +30 dBm to -40 dBm
  • RF acquisition length (Maximum RF acquisition time)
    • >2 GHz: 2.5 ms
    • >1 GHz - 2 GHz: 5 ms
    • >800 MHz - 1 GHz: 10 ms
    • >500 MHz - 800 MHz: 12.5 ms
    • >400 MHz - 500 MHz: 20 ms
    • >250 MHz - 400 MHz: 25 ms
    • >200 MHz - 250 MHz: 40 ms
    • >160 MHz - 200 MHz: 50 ms
    • >125 MHz - 160 MHz: 62.5 ms
    • <125 MHz: 79 ms
  • Trigger system
    • Main trigger modes: Auto, Normal, Single
    • Trigger coupling: DC, AC, HF reject (attenuates >50 kHz), LF reject (attenuates <50 kHz), noise reject (reduces sensitivity)
    • Trigger Holdoff Range: 20 ns to 8 s
    • Trigger modes: Edge, Sequence (B-trigger), Pulse Width, Timeout, Runt, Logic, Setup and Hold, Rise/Fall Time, Video
    • Optional trigger modes: Extended Video, I2C, SPI, USB, Ethernet, CAN, LIN, FlexRay, RS-232/422/485/UART, MIL-STD-1553, I2S/LJ/RJ/TDM, Parallel
  • Acquisition modes: Sample, Peak Detect, Averaging, Envelope, Hi Res, Roll
Type  Benchtop Portable  
Bandwidth  1.0 GHz 
Scope Channels  4  
Digital Channels  16  
Sample Rate  5.0 GSa/sec 
Memory Depth  Maximum record length: 20 Mpoints  
Status -
Launch Date Aug-2011